REED R8800 Specifications
- SOURCE & MEASUREMENT
Current Range: 0 to 19.99mA; 0 to 24mA
Resolution: 0.01mA; 0.1mA
Accuracy: ±(0.25% f.s. + 1 dgt.); ±(0.5% f.s. + 1 dgt.) - DC mV SOURCE
Voltage Range: -199.9mV to +199.9mV
Resolution: 0.1mV - Accuracy: ±(0.25% f.s. + 1 dgt.)
GENERAL
Display: LCD, max. 1999 counts
Power Supply: Single 9V battery
Dimensions: 150 x 70 x 40mm
Weight: 232 g
Sampling Time: Approx. 0.4 second
Operating Temperature: 0 to 50°C (32 to 122°F)
Operating Humidity: < 70% RH
REED R8800 Battery Replacement When the batteries become exhausted or drop below the operating voltage, “BAT” will appear in the left-hand side of the LCD display. 1. Disconnect the test leads from the meter. 2. Open the battery door by loosening the screw using a Phillips head screwdriver. 3. Insert the battery into battery holder, observing the correct polarity. 4. Put the battery door back in place. Secure with the two screws.